01 / Technology

Helium mass spectrometry, engineered for the line

02 / How it works
1
Tracer gas

Helium is inert, non-toxic and present in air at only ~5 ppm — background noise stays low, so even the smallest leak stands out.

2
Magnetic deflection

A 180° non-uniform magnetic field separates He ions (m/z 4, also ³He and H₂) from every other gas species in the analyzer.

3
Quantitative readout

The ion current maps linearly to leak rate — a calibrated number in Pa·m³/s, not a pass/fail guess. Verified against the built-in standard leak.

03 / Benchmark

Against imported mainstream models

ParameterQG-800 (8 L/s dry)Pfeiffer ASM 340 DINFICON UL3000 FabLeybold Phoenix Quadro Dry
Min. detectable leak, vacuum (Pa·m³/s)5.0×10⁻¹³5×10⁻¹³<5×10⁻¹³5×10⁻¹³
Min. detectable leak, sniffer (Pa·m³/s)≤ 5.0×10⁻⁹5×10⁻¹⁰<5×10⁻⁹1×10⁻¹⁰
He pumping speed at inlet (L/s)2.52.54>3.1
Dry forepump speed (m³/h)303.4>303
Response time≤ 0.5 s< 1 s< 1 s< 1 s
Startup time≤ 3 min~3 min< 3 min110 s
Ion sourceYttria-Ir dual filamentsYttria dual filamentsCoated Y-Ir dual filamentsLong-life Ir dual cathodes
InterfacesRS232 / IORS232 / IO / USB / fieldbus opt.USB / Ethernet / IOHDMI / USB / Ethernet / Wi-Fi
Weight98 kg45 kg~120 kg35 kg

Note: leak rates unified to Pa·m³/s (1×10⁻¹² mbar·L/s = 1×10⁻¹³ Pa·m³/s). Competitor figures compiled from public datasheets and manuals (retrieved 2026-09); refer to each vendor's official documentation for current data.

01
Sensitivity in the same class

Vacuum-mode detection floors are all in the 5×10⁻¹³ Pa·m³/s range — while our ≤ 0.5 s response reads faster.

02
9–10× the forepump speed

A 30 m³/h dry forepump matches the semiconductor-class INFICON and outruns the Pfeiffer / Leybold dry models by roughly an order of magnitude.

03
Ownership cost advantage

Domestic manufacturing means friendlier pricing, spares and service response. (Imported units are typically quoted around US$15k–45k depending on configuration.)

04 / Compliance & traceability

Calibrated, verifiable, auditable

Every QG-800 ships with a built-in channel-type standard leak (10⁻⁷ mbar·L/s) for daily verification. Methods follow GB/T 15823-2009 (helium leak testing practice) and GB/T 13979; calibration certificates are traceable to national standards.

GB/T 15823-2009
Helium mass-spectrometer leak testing practice
GB/T 13979
Mass spectrometer leak detector calibration
10⁻⁷ mbar·L/s
Built-in standard leak for daily checks
IQ / OQ
Qualification documentation for regulated industries
Want the raw numbers on your parts?

Free sample test — quantitative report, no commitment.